Introduction
The KWS-960C online turbidity sensor is designed and manufactured based on the principle of scattered light turbidity measurement. When a beam of light is projected into a water sample, the light is scattered due to the turbidity substances in the water sample. The turbidity in the water sample is calculated by measuring the intensity of the scattered light in the direction perpendicular to the incident light and comparing it with the internal calibration value. The final value is output after linearization processing.
Feature  Â
- 90° angle scattered light principle, built-in temperature sensor
- Support RS-485, Modbus/RTU protocol
- Optical fiber structure, strong resistance to external light interference
- Infrared LED light source with high stability
- IP68 protection, water depth within 20 meters
- Convenient, fast, stable and easy to maintain
Technical Specifications
Model | KWS-960 | |
Principle | Light scattering method | |
Range/Resolution | 0~20.00NTU | 0.01NTU/0.1℃ |
0~200.0NTU | 0.01NTU/0.1℃ | |
0~1000.0NTU | 0.1NTU/0.1℃ | |
Precision | ±3% or ±1.5NTU(0~20.00NTU)
±3% or ±2NTU(0~200.0NTU) ±5% or ±3NTU(0~1000.0NTU) |
|
Response Time (T90) | <60s | |
Lowest Detection Limit | 0.01 (0-20NTU) | |
0.3NTU | ||
Calibration Mode | Two-point calibration | |
Temperature Compensation | Automatic temperature compensation (Pt1000) | |
Output | RS-485(Modbus/RTU) | |
Working Conditions | 0~50℃,<0.2MPa | |
Storage Temperature | -5~65℃ | |
Installation | Immersion mount with 3/4NPT pipe thread | |
Cable Length | 5 meters, other lengths can be customized | |
Power Consumption | 0.2W@12V | |
Power Supply | 12~24VDC | |
Protection Grade | IP68 |